Portable Automatic Chemistry Analyzer

Specification for Chemistry analysis Machine:

ay Method  Reaction rate, end point, two-point, double wavelengths, coefficient, multiple standards, blank reagent, blank sample, absorbance at end point, QC, QC graph and turbidity
Temperature System PELTIER temperature control component, 25ºC, 30ºC, 37ºC, accuracy: ±1ºC
Light Source 6V 10W halogen lamp of more than 2000 hours lifespan
GML Precision ±2 nm
FWHM ≤10 nm
Stray Light ≤0.5(absorbance≥2.5)
Absorbance Range -0.214-3.000
Absorbance Linearity ±1%
Absorbance Repeatability ±0.001
Precision ≤1%
Carry Over ≤1%
Resolution 1mAbs/h
Stability ≤3mAbs/h
Flow Cell Permanent quartz flow cell, minimal volume is 32uL, and disposable flow cell is selectable
Sampling System Pressure auto releasing pump ensure the accuracy of suction volume
Absorption Volume 200uL-3000uL
Sample Volume 10uL-50uL
Display TFT LCD, touch screen
Working condition Temperature 10-30ºC, RH: 20-80%
Voltage  AC85-264V, auto adapt
Power ≤100W
Quality System CE